Operation
Using air temperature control
WARNING
Overheating of patient is possible if the extended set
value range is used.
Monitor the body temperature continuously.
If the set value exceeds the standard set value range
The following message appears in the upper section of the
screen.
»Confirm the extended range by pressing the rotary knob«.
Confirm the extended set value range = press the rotary
knob.
Continue increasing the set value = turn the rotary knob
clockwise.
The message »
>37.0
The following message appears in the upper section of the
screen.
»set value with rotary knob«
Confirm the set value = press the rotary knob.
The display returns to the standard screen. The measured
values are displayed.
The set value and »set: >37.0
alternately.
The yellow bar LED lights up.
If the yellow bar LED flashes, the set value has been
increased by more than 1.5
alarms are active.
If the specified set value is below the standard range
The following message appears in the upper section of the
screen.
»confirm extended range with rotary knob«.
Confirm the extended set value range = press the rotary
knob.
Continue decreasing the set value = turn the rotary knob
counter-clockwise.
46
o
C« appears on the screen.
« are displayed
o
C, see page 48, or other