Warming bed

The bed is enclosed by two side panels and two end panels
that can be opened independently of one another.
The side panels are equipped with holes for secretion and
drainage hoses.
The end panels are equipped with flexible grommets for hoses
and cables.
Fitting the panels
1 Insert the lower mounting pins of the panel hinge-pieces
into the guide slots on either side.
Ensure that the handle of the side panels faces outwards.
Position the panel semi-upright.
2 Press the panel down firmly until the pins lock into position
at the bottom of the slots.
Fold the panel up into the vertical position and allow
it to slide into the locking position.
Fit all four panels in this way.
Make sure that the panels are seated correctly.
To open:
Lift the panel all the way up until the upper pins
come out of their slots, then fold the panel down.
With lively infants or older children:
Use 230 mm high panels.

Fitting the inner panels

WARNING
The inner panels should be used with all patients,
whether the side panels are open or closed.
Otherwise the patient may fall out!
Insert the pins into the tapered holes in the corners of the
housing and press down slightly until the inner panel is
firmly in place.
Make sure that the panels are seated correctly.