6

How to Use the Vivo 3

For a summary description of the means of initiating and terminating the inspiratory phase
in each mode of the ventilator, please refer to .
6.1

Switch On the Vivo 3

Switching on an unpowered ventilator
Connect the power supply.
⇒The ventilator starts up and enters standby mode.
If the power supply is connected during the Power Failure alarm, the ventilator will start
the treatment directly instead of entering standby mode.
Switch on a powered ventilator
Press the On/Off button.
⇒The ventilator starts up and enters standby mode.
6.2

Start the Treatment

WARNING!
Risk of Asphyxia or Personal Injury
Using incorrect settings may cause personal injury or severe medical conditions, such as
hypercarbia, producing arterial acidemia.
The therapy settings shall be based on a physician's prescription. Changes to settings
must be made by authorized clinical personnel only.
It is possible to choose how the machine will start and stop. If you want to confirm the start-
ing and stopping or not, see 3.5.5.1 Device Settings, page 28
Confirm Start/Stop is set to On
Make sure that the Vivo 3 is in Standby mode.
1
Press and hold the On/Off button until the
Starting Treatment progress bar is filled.
⇒The Vivo 3 now performs a function test indicated
by a short beep and then starts the treatment. Check
that the self test is performed successfully, this is
indicated by:
• A short beep indicating successful alarm
• The alarm LED first lights yellow, then red
• The audio pause LED lights yellow
• In about a second, both LEDs are turned off
If the function test beep absents, take the Vivo 3 out of use and contact your supplier of the
Vivo 3 .
40 How to Use the Vivo 3
User Manual
signalling