Alarm Reset; Alarm And Warning Descriptions - Breas Clearway 2 User Manual

Hide thumbs Also See for Clearway 2:
Table of Contents

Advertisement

Alarm Reset

An alarm condition will be automatically reset when the cause of the alarm has
been corrected or the alarm condition no longer exists.
If an alarm cannot be corrected, stop using the Clearway 2 and
request technical support
When an alarm has been acknowledged on screen, the alarm message will
disappear unless the fault is reoccurring or still exists. If the fault still exists
after acknowledgment, the alarm symbol will remain on screen and enables the
user to view the current list of live alarms.

Alarm and Warning descriptions

High Pressure Alarm
Item
Definition
Priority
Modes
Possible Cause
Clearway 2 action
On Screen Message
DOC-001000 Ver 6
Description
120% of set pressure is detected for > 2 seconds
Medium
All
Cough during inspiration
Treatment stops. The on-screen message must be
acknowledged. Treatment cannot be started until
message has been acknowledged
"High Pressure Alarm!
Treatment has stopped!
- Restart the Clearway 2
- Contact your supplier if fault persists"
Alarms
Clearway 2 User Manual
50
.