4.1.6. Clean the sample tube adapter

If sample has spilt into the adapter.
As needed
Switch OFF .
Always wear protective garments and
gloves when handling the waste fluid.
After work is completed, wash hands
1
with disinfectant.
Switch OFF .
Remove adapter and clean it.
2
Remove adapter and clean it.
- Clean with distilled water
or
- Clean with CELLCLEAN diluted 10 times or
more.
Rinse the adapter with distilled water after-
wards

4.1.7. Perform auto rinse

Clean tubings and drain waste.
Even if the message "Auto Rinse" is not displayed, auto rinse can be executed by pressing
"Menu", "Maint." and then "Auto Rinse".
As needed
Biological risks
1
2
3
Press "Execute".
Press "Execute".
Process starts.
Process starts. Afterwards "Background
check" will be performed automatically.
Background check result will ap-
pear.
Press "Result". Background check result will
appear.
i
4 - 6
pocH-100
Revised June 2008