4.2. TECHNICAL MAINTENANCE-Qualified personnel only!
4.2.1. Clean aperture of TD chamber
If the clogging of the aperture cannot be removed by the automated sequence, the aperture must be cleaned manually with the transducer brush.
As needed
Press "Menu".
1
Press "Menu".
Press "Maint.".
2
Press "Maint.".
Press "Drain T D Chamber".
3
Press "Drain T D Chamber".
4 -13
i
pocH-100
Revised June 2008
Press "Execute".
4
7
Press "Execute".
Process starts.
5
8
Process starts.
The screen of process completed.
6
9
The screen of process completed.
Switch OFF .
10
Switch OFF .
Disconnect power cord.
11
Disconnect power cord to avoid electrical
shock.
Unscrew cover.
12
Unscrew cover by using the supplied screw-
driver.
Open side cover and open detec-
tor cover.
Open side cover and open detector cover.
Check if fluid has not been drained.
Check if fluid has been drained. If not, re-
move it with the pipette.
Remove the plug.
Place cloth under plug and remove plug.