1.3. SETUP

1.3.5. Barcode reader (optional)

Scans the barcode on the sample collection tube and automatically enters the sample number into the system.
Plug cable into "BR" connection.
1
Plug cable into "BR" connection.
Turn the main switch ON, press
the trigger switch and check red
LED.
2
Turn the main switch ON, press the trigger
switch and check red LED.
ID Barcode specifications
Barcode labels' specifications must match the requirements for the barcode reader.
Symbol
Check digit
None
ITF
Modulus 10
None
NW-7
Modulus 1 1-16
None
CODE 39
Modulus 43
CODE 128
Modulus 103
JAN-8
Modulus 10
JAN-13
Modulus 10
1- 13
1- 13
i
pocH-100
Revised April 2013
Number of digits
Max. 14 digits (sample ID)
Max. 14 digits (sample ID) + 1 digit (check digit) = Max. 15 digits
Max. 15 digits (sample ID)
Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
Max. 15 digits (sample ID)
Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
Max. 7 digits (sample ID) + 1 digit (check digit) = Max. 8 digits
Max. 12 digits (sample ID) + 1 digit (check digit) = Max. 13 digits
Important
• Disconnect the instrument's mains before con-
necting the peripheral device. Else, the two de-
vices may not interchange any data.
• In identifying patient samples, maximum integrity
of data is required.
To avoid barcode identification mistakes, use a
check digit whenever the sample barcode is used.
If a check digit is not used, the chances of the bar-
code being misread increase.
Note
The barcode reader is not included as a standard
accessory. Refer to the barcode reader manual
for detailed information on the connection of the
barcode reader.