SYSMEX pocH-100i Instructions For Use Manual page 94

Automated hematology analyzer
Table of Contents

Advertisement

6.3. ERROR MESSAGES, POSSIBLE CAUSES AND ACTIONS TO RESOLVE THE ERROR
R-Cover open
Possible cause
Action to resolve the error
Special notes, if error re-
mains
PLT Noise Error
RBC Noise Error
WBC Noise Error
Possible cause
Action to resolve the error
Special notes, if error re-
mains
PPMC Cont. Error
Possible cause
Action to resolve the error
Special notes, if error re-
mains
6 - 7
i
pocH-100
Revised February 2011
The right side cover was opened.
After closing the right side cover, press the "Execute" button
to clear the error, and the recovery sequence is executed.
If the recovery process is completed correctly, then Auto
Rinse and background check are performed in sequence, and
the screen returns to the Main screen.
Not ready until the error is resolved.
• Effects of external electric noise interference. Control board
malfunction.
• Hydraulic line or aperture partly clogged.
Separate the source of electric noise from the instrument.
Ready for next analysis, however, the following sample re-
sults may be incorrect.
Error in the motor controller circuit.
Turn the main switch OFF , then turn it ON again.
Not ready until the error is resolved.
HC Buffer Full
LAN Buffer Full
Print Buffer Full
The amount of data for output or printout is too large to be
Possible cause
processed.
Action to resolve the error
Press "Back" to clear the error.
Special notes, if error re-
Not ready until the error is resolved.
mains
No Printer Paper
Possible cause
No printer paper in the built-in thermal printer.
• After refilling paper in the built-in thermal printer, press "Re-
try" to clear the error.
The data currently being printed will restart printing from the
Action to resolve the error
beginning of that sample.
• After refilling paper in the built-in thermal printer, press
"Cancel" to cancel the data being printed.
Also discard the waiting data and the error is cleared.
Special notes, if error re-
Not ready until the error is resolved.
mains
Printer Error
• The built-in thermal printer has an error.
Possible cause
• The printer cover is open.
• Check if the paper is set correctly and close the cover (see
page 4-1 1).
• After checking the built-in thermal printer, press "Retry" but-
ton to clear the error.
Action to resolve the error
The data currently being printed will restart printing from the
beginning of that sample.
• After checking the built-in thermal printer, press "Cancel" but-
ton to cancel the data being printed.
Also discard the waiting data and the error is cleared.
Special notes, if error re-
Not ready until the error is resolved.
mains