3.2. PERFORMING QUALITY CONTROL
3.2.3. STEP 3: a) Performing quality control: L-J method (factory setting) (for close piercer)
Daily
Press the
-button or press "Operator".
1
Press the
-button or press "Operator".
Select a operator ID.
Operator 1
Operator 2
2a
Operator 3
more
Select a operator ID.
or
Enter the operator ID and press "Ent.".
2b
A
Enter the operator ID and press "Ent.".
3 - 9
i
pocH-100
Revised February 2011
Press "QC".
3
Press "QC".
Select analyzing File.
4
1/2
close
Select analyzing File.
Open sample position.
5
Open sample position.
Insert adapter.
6
Insert correct adapter.
Insert control blood and close
the door.
7
10
Insert control blood and close door.
Press "Run".
8
11
Press "Run".
The analysis starts.
9
The analyzing screen will appear.
Press "
" or "
" to scroll
screens.
Press "
" or "
" to scroll the results
screens.
Press "Quit".
Press "Quit" to accept analysis results.
Depending on QC settings, results will be
printed/sent to host computer automatically.