Clean aperture gently.
13
Apply CELLCLEAN to transducer brush,
clean aperture gently.
Close all covers. Tighten the screws.
14
Close all covers. T ighten the screws.
Switch ON the instrument.
15
*
Plug in power cord Switch ON instrument.
(
1)
Biological risks
Always wear protective garments and gloves for
all service and maintenance work. Use only speci-
fied tools and parts. After completion of work,
wash hands with disinfectant. Instruments that got
in contact with blood have infectious potential.
Danger, electric shock
Re-attach the transducer chamber plug securely.
Otherwise, reagent leakage might occur, which
may cause a short circuit or electrical shock.
Important
If performing maintenance work, use only the
specified tools and parts. Install only such spare or
*
replacement parts intended for this equipment.
1
. Be sure to turn the main switch ON after com-
pletion of the cleaning procedure.
If the auto rinsing is not performed after clean-
ing, the instrument may malfunction.
Note
Before storing, rinse the brush with water to com-
pletely remove the CELLCLEAN.
i
pocH-100
Revised June 2008
4 -14