Advertisement

4.1. CLEANING

4.1.5. Clean waste chamber

Even if the message "Clean waste chamber" is not displayed, this maintenance can be executed by pressing "Menu", "Maint." and then "Clean waste chamber".
Every 3 months or 1500 samples
Cleaning instruction message is
displayed.
1
Cleaning instruction message is displayed.
Add 3mL of CELLCLEAN into the
sample tube.
2
Add 3mL of CELLCLEAN into empty sample
tube.
Open sample position.
3
Open sample compartment.
4 - 5
i
pocH-100
Revised February 2011
Insert adapter.
4
7
Insert adapter.
Set CELLCLEAN in place, close
compartment.
5
8
Set CELLCLEAN in place, close compartment.
* For open pipette, remove a cap.
Press "Execute".
6
Press "Execute".
Process begins.
• A message will appear if either the counter value
exceeds 1500 or 3 months have passed since the
last "Clean waste chamber".
• After cleaning, the counter resets automatically.
Process begins. Afterwards "Auto Rinse"
and "Background check" will be performed
automatically.
Cleaning completed. Remove
CELLCLEAN.
When this screen is displayed, remove
CELLCLEAN.
Note