Performing Quality Control: X Method (For Open Pipette) - SYSMEX pocH-100i Instructions For Use Manual

Automated hematology analyzer
Table of Contents

Advertisement

3.2. PERFORMING QUALITY CONTROL
3.2.3. STEP 3: b) Performing quality control: X
Daily
Press "QC".
1
Press "QC".
Select analyzing File.
2
Select analyzing File.
Open sample position.
3
Open sample position.
The sample tube caps have to be removed
before measurement.
3-15
i
pocH-100
Revised February 2011
_
method (for open pipette)
Set the spacer to an adapter.
4
Set the spacer to an adapter.
Insert adapter.
5
Insert correct adapter.
Remove the cap.
6
Remove the cap.
Insert control blood and close
the door.
7
10
Insert control blood and close door.
Press "Run".
8
11
Press "Run".
Press "Next" to perform the sec-
ond analysis.
9
12
The first analysis results appear. Press
"Next" to perform the second analysis.
Mix well and insert it back into
the sample position.
Remove the control blood, mix well and
insert it back into the sample position.
Press "Run".
Press "Run" to perform second analysis.
The final results appear.
The final results appear. Scroll the results
by pressing " " or "
" and press "Print".