Sample Preparation; Instrument Preparation - SYSMEX BK265455 Operator's Quick Manual

Automated hematology analyzer clia waived
Table of Contents

Advertisement

XW-100 SAMPLE PROCESSING

Sample Preparation

The blood sample must:
Be collected in an EDTA-K2 or K3 (purple top) vacuum tube as shown on the instrument's display screen.
Be run within 8 hours of having been drawn, if stored at room temperature, or within 36 hours if
refrigerated.
Required sample volume: > 1 mL

Instrument Preparation

If the screen displays "QC Required," refer to the Quality Control section of this Quick Guide.
If the screen displays "General Care Required," refer to the General Care section of this Quick Guide.
(Continued on the next page)
4