4.2. TECHNICAL MAINTENANCE-Qualified personnel only!

4.2.2. System fuse replacement

As needed
Switch OFF .
1
Switch OFF .
Unplug power cable.
2
Unplug power cable.
Turn fuse holder cap counter-
clockwise.
3
Turn fuse holder cap counter-clockwise with
screwdriver.
4 -15
i
pocH-100
Revised June 2008
Replace the fuse(s).
4
Remove fuse holder cap. Replace fuse(s). Put
fuse holder cap back in place.
Switch ON.
5
Plug in power cable. Switch ON.
Danger, electric shock
To avoid electrical shock, disconnect power supply
before servicing.
Important
Replace only with fuse of the specified type and
correct current specifications.