Performing Quality Control; A) Settings For Control Blood Information (Quality Control File) - SYSMEX pocH-100i Instructions For Use Manual

Automated hematology analyzer
Table of Contents

Advertisement

3.2. PERFORMING QUALITY CONTROL

3.2.1. STEP 1: a) Settings for control blood information (quality control file)
Press "Menu".
1
Press "Menu".
Press "QC".
2
Press "QC".
Press "Setup".
3
Press "Setup".
3 - 5
i
pocH-100
Revised June 2008
Enter "Lot ID" and "Expire", then
press " ".
4
7
Enter "Lot ID" and "Expire" manually or by
barcode reader. Press "
".
Enter TARGET and LIMIT value
for each parameter.
5
8
Enter TARGET and LIMIT value for each pa-
rameter manually or by barcode reader.
Switch to remaining 5 QC file
setting screens.
6
Press "
" to switch to remaining 5 QC file
setting screens.
Press "Save".
The operator must ensure the control blood is not
beyond expiration date. The analyzer does not
check the expiration dates entered in the QC file.
• Before using new control blood, enter the control
blood information: Control blood lot ID and Ex-
After completion of entering all parameters,
piration date. For each control parameter enter:
press "Save".
Target value and Target Limit
• Control blood must always be stored according
to the instructions in the package insert.
Press "OK".
Changes in the analysis results over time must
be recorded either by printing results or by other
means such as storing the data on PC.
Press "OK".
Incorrect analysis results
Important
Note