XW™-100 AUTOMATED HEMATOLOGY ANALYZER
CLIA WAIVED
OPERATOR'S QUICK GUIDE
Code No. BK265455 en-am
PRINTED IN JAPAN
Date of Last Revision: 12/2022
Document Version: 3.0
© SYSMEX CORPORATION
Software Version: Ver.1

Advertisement

Table of Contents
loading

Summary of Contents for SYSMEX BK265455

  • Page 1 B E YO N D A B E T T ER B OX ™ XW™-100 AUTOMATED HEMATOLOGY ANALYZER CLIA WAIVED OPERATOR'S QUICK GUIDE Code No. BK265455 en-am PRINTED IN JAPAN Date of Last Revision: 12/2022 Document Version: 3.0 © SYSMEX CORPORATION...
  • Page 2: Daily Cleaning And Disinfecting

    XW-100 GENERAL CARE Daily Cleaning and Disinfecting NOTE: The following must be cleaned and disinfected every day that patient samples are tested to prevent the spread of infection: • XW-100 cover and display • Exterior of the sample chamber • White sample adapter used for testing Put on personal protective equipment (PPE);...
  • Page 3: Quality Control

    We run QC CHECK on the XW-100 to ensure the system is performing correctly. • Sysmex will ship XW QC CHECK every 28 days. Upon arrival, unpack and store the container at room temperature.
  • Page 4: Sample Preparation

    XW-100 SAMPLE PROCESSING Sample Preparation The blood sample must: • Be collected in an EDTA-K2 or K3 (purple top) vacuum tube as shown on the instrument's display screen. • Be run within 8 hours of having been drawn, if stored at room temperature, or within 36 hours if refrigerated.
  • Page 5: Sample Processing

    XW-100 SAMPLE PROCESSING Sample Processing Enter the Patient ID. 6. Confirm the Patient ID. Confirm purple top blood collection tube is properly labeled. NOTE: This is the only sample tube with a purple top that may be used. WARNING/LIMITATIONS: - Samples from children under 2 years of age must not be tested.
  • Page 6: Review Results

    XW-100 SAMPLE PROCESSING Sample Processing , continued from previous page 10. Follow the on-screen instructions to mix the 12. Wait for the test cycle to complete. collection tube. 11. Insert the collection tube, then close the door. Before inserting the tube: After inserting the tube: Review Results IMPORTANT:...
  • Page 7 XW-100 SYSTEM COMPONENTS XW-100 Power Cord Ethernet Cable Waste Tubing Barcode Reader Container Spout Kit Reagent Tray Green Adapter White Adapter Thermal Paper (XW pack D, XW pack L) Waste Bottle XW CELLCLEAN XW QC CHECK XW pack D XW pack L...
  • Page 8 XW-100 INSTRUMENT SHUTDOWN Shutdown of the XW-100 is not required. If your laboratory would like to shutdown the instrument, please refer to the Shutdown Quick Guide. NOTE: Once shutdown has been performed, the following will need to be completed upon start up: The internet connection will need to be reestablished.
  • Page 9: Waste Disposal

    XW-100 SAFETY INFORMATION Waste Disposal BIOLOGICAL RISKS: Once used, this instrument and its accessories are considered biohazardous. Do dispose of the instrument, components, or waste in public trash containers. Please contact your Technical Assistant Center (TAC) for instructions on instrument, component, and waste disposal.1-800-779-7639 EMC Information •...
  • Page 10: Instrument Failure

    XW-100 SAFETY INFORMATION Instrument Failure If the instrument fails, contact your Technical Assistant Center (TAC). Please note the ERROR CODE to help the representative provide quick assistance. 1-800-779-7639 Equipment Operation and Ethernet Cable Connections The connection of a non-shielded equipment interface cable to this equipment will invalidate the IEC Certification of this device and may cause interference levels which exceed the limits established by the IEC for this equipment.
  • Page 11 NOTES...
  • Page 12 Sysmex Corporation Sysmex America, Inc. 577 Aptakisic Road 1-5-1 Wakinohama-Kaigandori Chuo-ku, Kobe 651-0073, Japan Lincolnshire, IL 60069, U.S.A. Tel. +1 (800) 379-7639 www.sysmex.com/us Program availability varies by location. Programs and specification subject to change without notice.