SYSMEX pocH-100i Instructions For Use Manual page 35

Automated hematology analyzer
Table of Contents

Advertisement

Select a operator ID .
12a
Select a operator ID .
or
Enter the operator ID and press "Ent.".
12b
Enter the operator ID and press "Ent.".
Open sample position and insert
the adapter.
13
Open sample position and insert the adapter.
Remove the cap, set sample into
the adapter and close the door.
14
Remove the cap, set sample into the adapter
and close the door.
Press "RUN".
15
Press "RUN" and confirm warning with "OK".
The analysis starts.
16
The analysis starts. The results are displayed
within approx. 90-125 seconds and are stored
automatically.
Analysis results are automatical-
ly printed according to settings.
17
Analysis results are automatically printed/
sent to host computer according to settings
(from page 5-2 ).
Incorrect analysis results
• All performance data cited in this manual were
generated using specimens in EDTA anticoagulant.
Results may differ using other anticoagulants.
• To prevent aggregation, the sample should be
diluted and analyzed immediately after collection,
else correct analysis results will not be obtained.
Important
• Do not open the sample position while aspirating!
• When using micro tubes, place them in the cor-
rect adapter and make sure the cap is removed.
Note
• The sample ID can set be up to 15 characters. If a
sample's ID is set to „0" the result for this sam-
ple will not be stored.
• In the pre-diluted mode only the 8 CBC param-
eters are displayed.
• When analyzing in the pre-diluted mode, the his-
togram curves, distribution analysis data and the
distribution flags are not displayed.
The titles and axes of the Histograms are dis-
played.
i
pocH-100
Revised February 2011
2-10