Automatic Calibration - SYSMEX pocH-100i Instructions For Use Manual

Automated hematology analyzer
Table of Contents

Advertisement

5.2.2. Automatic calibration

The instrument determines the calibration value automatically by analyzing 5 calibration samples.
Press "Menu".
1
Press "Menu".
Press "Calib.".
2
Press "Calib.".
Press "Auto Calib.".
3
Press "Auto Calib.".
Enter reference values and press
"Ent.".
4
Enter reference values for 5 each sample
and press "Ent.".
Press "Next" and "OK".
5
Press "Next" and "OK".
Set first sample and press "Run".
6
Set first sample and press "Run" and repeat
this for the remaining 4 samples.
Press "Quit.".
• Automatic calibration is always performed in the
WB mode.
• It is important to analyze the samples according to
their reference values. The values of the sample to
7
be analyzed are indicated by the underline cursor.
• Set "0" for any parameter which does not need to
be calibrated.
• Do not open the sample position while "Aspirating"
is displayed. Only open it when "Running" is dis-
After completion of all analyses, press
played.
"Quit".
Press "OK".
If an error occurs during analysis, press "Cancel" and
analyze again.
8
Press "OK".
Important
Note
i
pocH-100
Revised June 2008
5 -12