4.1. CLEANING
4.1.3. Shutdown
Removes deposits in the instruments tubing.
Daily
Press "Shutdown".
1
Press "Shutdown".
Press "Execute".
2
Press "Execute".
Process begins.
3
Process begins.
4 - 3
i
pocH-100
Revised June 2008
Follow instructions on the screen.
Then switch OFF.
4
Follow instructions on the screen. Then
switch OFF .
Incorrect analysis results
• Deposits in the instrument's tubing can cause
incorrect analysis results. Therefore, the trans-
ducer chambers and diluted sample tubes must
be cleaned.
• If the instrument is used continuously without
performing the shutdown sequence, protein clot-
ting may cause incorrect analysis results or it
may damage the instrument.