2.1. ANALYSIS
2.1.5. b) Analysis in whole blood (WB) mode (for open pipette)
• Check blood with added EDTA anticoagulant, aspirated sample volume ~15μL.
• Required sample volume: >100μL for the ø 13mm tube, > 50μL for the micro tube.
Press "Sample ID".
1
Press "Sample ID".
Enter the ID and press "Ent.".
2
Enter the ID manually or by barcode reader
and press "Ent.".
Open sample position.
3
Open sample position.
The sample tube caps have to be removed
before measurement.
2 - 7
pocH-100i Revised February 2011
Set the spacer to an adapter.
4
7
Set the spacer to an adapter.
Insert adapter.
5
8
Insert the correct adapter.
Mix the sample.
6
9
Gently mix the sample.
Remove the cap.
10
Remove the cap.
Insert the sample tube and close
the door.
11
*
Insert sample tube and close the door. (
1)
Press "RUN".
Press "RUN".
The analysis starts.
The analysis starts. The results are displayed
within approx. 90-125 seconds and are
stored automatically.
Analysis results are automatical-
ly printed according to settings.
Analysis results are automatically printed/
sent to host computer according to settings
(from page 5-2).