Remove Clog From Transducer Aperture - SYSMEX pocH-100i Instructions For Use Manual

Automated hematology analyzer
Table of Contents

Advertisement

4.1. CLEANING

4.1.8. Remove clog from transducer aperture

If an "Aperture Clog" error occurs, the clog must be removed.
Even if the message "Aperture Clog" is not displayed, automatic clog removal can be executed by pressing "Menu", "Maint." and then "Clog Removal".
As needed
Press "Execute".
1
*
Press "Execute";automatic clog removal
starts. (
1)
4 - 7
i
pocH-100
Revised June 2008
*
Note
1
. To return to the main screen with-
out performing automatic clog re-
moval, press "Cancel".