Waters ACQUITY UPLC Overview And Maintenance Manual page 96

Sample manager flow through needle
Table of Contents

Advertisement

Figure 3–50: Bending needle tubing at needle-tubing clamp
1
Needle-tubing clamp
Needle tubing
Rounded tool
25. Reinstall the access panel and use the T20 TORX driver to tighten the two screws that
secure the panel to the front of the unit.
Requirement:
Ensure that the seal-port tube and the sample-needle tubing are routed
through the gap in the access panel.
Figure 3–51: Tubing routed through access-panel gap
1
2
3
Seal-port tube
Access-panel gap
Sample-needle tubing
2
3
June 11, 2019, 715005051 Rev. D
Page 96