7.
Remove the wash tubing from the clip on the inside wall of the sample compartment.
Figure 3–9: Wash tubing secured by clip on sample-compartment wall
1
2
Wash tubing
Clip
Notice:
To avoid damaging the adapter fitting, ensure that you hold it while
unscrewing the wash-tube fitting.
8.
While holding the adapter fitting, unscrew the finger-tight wash-tube fitting from the adapter
fitting.
Figure 3–10: Wash-tube and adapter fittings
1
Adapter fitting
Wash-tube fitting
9.
Unscrew the finger-tight adapter fitting from the seal assembly.
10. Slide the support sleeve from the housing and guide the seal-port tube through the slots.
Requirement:
Hold the assembly in a vertical position when disassembling it. Doing so
helps ensure that the spring remains in the proper position.
2
June 11, 2019, 715005051 Rev. D