Waters ACQUITY UPLC Overview And Maintenance Manual page 71

Sample manager flow through needle
Table of Contents

Advertisement

7.
Remove the wash tubing from the clip on the inside wall of the sample compartment.
Figure 3–9: Wash tubing secured by clip on sample-compartment wall
1
2
Wash tubing
Clip
Notice:
To avoid damaging the adapter fitting, ensure that you hold it while
unscrewing the wash-tube fitting.
8.
While holding the adapter fitting, unscrew the finger-tight wash-tube fitting from the adapter
fitting.
Figure 3–10: Wash-tube and adapter fittings
1
Adapter fitting
Wash-tube fitting
9.
Unscrew the finger-tight adapter fitting from the seal assembly.
10. Slide the support sleeve from the housing and guide the seal-port tube through the slots.
Requirement:
Hold the assembly in a vertical position when disassembling it. Doing so
helps ensure that the spring remains in the proper position.
2
June 11, 2019, 715005051 Rev. D
Page 71