12. Use the 1/4-inch open-end wrench to reattach the fittings associated with the tubing for the
sample needle and the seal-port tube, and tighten them as much as 1/6-turn beyond finger-
tight for existing fittings, or 3/4-turn beyond finger-tight, for new fittings.
13. Reattach all finger-tight fittings and plugs, ensuring that the tubes bottom in their fitting
holes.
14. Close the fluidics compartment door.
15. Power-on the sample manager.

3.13 Cleaning the exterior of the equipment

Warning:
• ensure that the electrical power to the equipment is interrupted;
• when cleaning the surface of the equipment, apply water to a cloth, and then wipe
the instrument or device. Do not spray or otherwise apply water directly onto any
equipment surface.
Warning:
process.
Required tools and materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
To clean the exterior of the equipment:
Clean surfaces of the equipment using only a clean, soft, lint-free paper towel or clean cloth
dampened with water.
To avoid electric shock,
To avoid personal injury, use eye and hand protection during the cleaning