4.
Grasp the leak sensor by its serrations and pull upward on it to remove it from its reservoir.
Figure 3–2: Leak sensor serrations
1
Serrations
5.
Unpack the new leak sensor.
6.
Align the leak sensor T-bar with the slot in the side of the leak sensor reservoir and slide
the leak sensor into place.
Figure 3–3: Aligning lower leak sensor T-bar with slot
T-bar
Slot in leak sensor reservoir
7.
Connect the leak sensor connector to the front of the sample manager.
Figure 3–4: Attaching leak sensor connector
June 11, 2019, 715005051 Rev. D
Page 67