Waters ACQUITY UPLC Overview And Maintenance Manual page 67

Sample manager flow through needle
Table of Contents

Advertisement

4.
Grasp the leak sensor by its serrations and pull upward on it to remove it from its reservoir.
Figure 3–2: Leak sensor serrations
1
Serrations
5.
Unpack the new leak sensor.
6.
Align the leak sensor T-bar with the slot in the side of the leak sensor reservoir and slide
the leak sensor into place.
Figure 3–3: Aligning lower leak sensor T-bar with slot
T-bar
Slot in leak sensor reservoir
7.
Connect the leak sensor connector to the front of the sample manager.
Figure 3–4: Attaching leak sensor connector
June 11, 2019, 715005051 Rev. D
Page 67