Figure 3–23: Adapter fitting in support sleeve
1
Adapter fitting
Support sleeve
24. Ensure that the seal-port tube is bent upward, at a 90-degree angle, and that it is
approximately 2.5 cm (1.0 inch) from the housing.
Note:
To avoid binding the seal-port tube, ensure that the tube does not contact either
side of the inject/wash station assembly.
Figure 3–24: Seal-port tube bent upward at right angle
1
Housing
Seal-port tube bent upward
25. While holding the adapter fitting, screw the wash-tube fitting into the adapter fitting.
2
2
June 11, 2019, 715005051 Rev. D
Page 79