Attribute
Overvoltage category
Pollution degree
Moisture protection
Line voltages, nominal
Voltage range
Frequency
Maximum power draw
a. Protection Class I – The insulating scheme used in the instrument to protect from electrical shock. Class I identifies a single
level of insulation between live parts (wires) and exposed conductive parts (metal panels), in which the exposed conductive
parts are connected to a grounding system. In turn, this grounding system is connected to the third pin (ground pin) on the
electrical power cord plug.
b. Overvoltage Category II – Pertains to instruments that receive their electrical power from a local level such as an electrical
wall outlet.
c.
Pollution Degree 2 – A measure of pollution on electrical circuits that can produce a reduction of dielectric strength or
surface resistivity. Degree 2 refers only to normally nonconductive pollution. Occasionally, however, expect a temporary
conductivity caused by condensation.
d. Moisture Protection – Normal (IPX0) – IPX0 means that no Ingress Protection against any type of dripping or sprayed water
exists. The "X" is a placeholder that identifies protection against dust, if applicable.
B.4 SM-FTN and bioSM-FTN input/output specifications
Attribute
Event output relay (Inject Start Out)
Digital input signal (Inject Hold In)
B.5 SM-FTN and bioSM-FTN performance specifications
Item
Injection volume
range
b
c
d
Specification
• With the standard loop fitted, 0.1 to 10.0 µL, in 0.1 µL increments.
• With one of the optional extension loops fitted, as much as the volume
of the extension loop (50, 100, 250, or 1000 µL).
June 11, 2019, 715005051 Rev. D
Specification
II
2
Normal (IPXO)
Grounded AC
100 to 240 Vac
50 to 60 Hz
400 VA
Specification
Maximum current: 0.5 A
Maximum input voltage: 30 Vdc
Contact resistance: 0.2 ohms
Maximum input voltage: 30 Vdc
Logic high: 3.0 Vdc
Logic low: 1.9 Vdc
Page 120