3.9 Replacing the sample needle and needle guide

Warning:
To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning:
Required tools and materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• Needle-nose pliers
• 1/4-inch open-end wrench
• T6 TORX driver
• T10 TORX driver
• T20 TORX driver
• Needle assembly kit
To replace the sample needle and needle guide:
1.
Power-on the sample manager.
2.
Remove any sample plates from the sample chamber.
3.
In the console, select the sample manager, and then click Maintain > Replace
components > Needle.
Result:
4.
Click OK.
Result:
accessible position.
5.
Open the sample compartment door and the fluidics compartment door.
6.
Using a T20 TORX driver, loosen the two captive screws that secure the access panel, and
then remove the panel.
To avoid eye injury, use eye protection when performing this procedure.
A warning for the user to keep clear of the chamber appears.
Doing so invokes a wizard that moves the sample-needle assembly to an
June 11, 2019, 715005051 Rev. D
Page 84