Waters ACQUITY UPLC Overview And Maintenance Manual page 95

Sample manager flow through needle
Table of Contents

Advertisement

Figure 3–49: Sample-needle assembly installed in sample compartment
7
6
5
Mounting sleeve in needle-tubing clamp
Needle-mounting cylinder in mounting cavity
Needle tubing in needle guide
Needle tip
Needle tubing
Lower RZZ mechanical rod
Needle tubing in notches
23. Using the T10 TORX driver, tighten the screw on the needle-tubing clamp.
Note:
You might need to slightly rotate or twist the needle to ensure that it routes straight
down the channel before tightening the screw on the needle-tubing clamp.
24. At the needle-tubing clamp, bend the needle to the right, toward port 4 on the injection
valve.
Tip:
Use a small, rounded tool (such as a T10 TORX driver) to aid in bending the tubing.
June 11, 2019, 715005051 Rev. D
Page 95
1
2
3
4