Waters ACQUITY UPLC Overview And Maintenance Manual page 122

Sample manager flow through needle
Table of Contents

Advertisement

Item
Specification
Maximum sample
768 in two, 384-well Waters-certified plates, or 96 in 2-mL vial holders.
capacity
Four additional positions for dilution functions.
For more information, see the Waters Sample Vials and Accessories
Brochure, or visit the
Sample
Between 4 and 40 °C, settable in increments of 0.1 °C; maintains 19 °C
compartment
below ambient with a tolerance range of between -2.0 and +4.0 °C
temperature range
• At a set point of 4 °C with ambient temperature <23 °C and humidity
• At ambient temperatures >23°C and humidity <80%, maintains an
Minimum sample
See the graph that follows this table for the achievable sample
compartment
compartment temperature and expected variation at various sample
temperature
temperatures.
specifications
The graph shows achievable sample compartment temperature and
expected variation at various sample temperatures.
Recommended
See the diagram that follows this table for the recommended locations for
locations for
temperature sensors on the sample tray when validating specifications.
temperature
sensors
Temperature
±0.5 °C at sensor
accuracy
Temperature
±1.0 °C (at the sensor with sample compartment door closed)
stability
Injection needle
Integral, active, and programmable
wash
Minimum sample
3 µL, residual, using total recovery 2-mL vials (zero offset)
required
plate selector
<80%, maintains a sample temperature of 2 to 8 °C.
average sample temperature of 18 °C below ambient, ±3.0 °C.
June 11, 2019, 715005051 Rev. D
Page 122
and
vial selector
on the Waters website.