Figure 3–25: Wash-tube fitting in adapter fitting
1
Adapter fitting
Wash-tube fitting
26. Place the inject/wash station assembly in its original location on the sample-compartment
floor and align the screw hole with the hole in the floor, ensuring that the wash drain of the
injection port is in the needle-wash basin.
Tip:
There are unused holes in the sample-compartment floor.
27. Adjust the amount of tubing in the sample compartment by sliding it in and out of the foam.
Requirement:
The wash tubing is secured to the wall and must not interfere with
operation of the sample tray or the vertical motion of the wash port.
28. Route the wash tubing through the clip on the inside of the sample compartment.
Figure 3–26: Wash tubing secured by clip on sample-compartment wall
1
2
Wash tubing
Clip
29. Using the T10 TORX driver, tighten the screw that secures the seal assembly to the
sample-compartment floor.
2
June 11, 2019, 715005051 Rev. D
Page 80

Hide quick links: