4.
Using the T20 TORX driver, loosen the two captive screws that secure the access panel,
and then remove the panel.
Figure 3–5: Captive screws on access panel
Captive screws (2)
5.
Using the 1/4-inch open-end wrench, unscrew the seal-port tube's fitting from port 1 of the
injection valve.
Figure 3–6: Location of seal-port tube on injection valve
Seal-port tube
6.
Using the T10 TORX driver, remove the screw that secures the inject/wash station
assembly to the sample-compartment floor and lift the assembly upward, out of its
mounting hole.
1
1
June 11, 2019, 715005051 Rev. D
Page 69