Figure 2–13: Aligning leak sensor T-bar with slot
Leak sensor T-bar
Slot
5.
Attach the leak sensor connector to the upper port on the front of the device.
Figure 2–14: Attaching leak sensor connector
Leak sensor connector
Leak sensor port on front of device
6.
Power-on the sample manager.
7.
In the console, select Sample Manager > Control > Reset SM to reset the sample
manager.
8.
In the console, enable the leak sensor (or sensors) to activate the leak-detection capability
(see the console online Help).

2.10 Calibrating the needle

You must calibrate the needle before you use the sample manager for the first time and
whenever you replace the sample needle. Failing to calibrate the needle can damage it. The
calibration procedure is identical for all needles.