Waters ACQUITY UPLC Overview And Maintenance Manual page 32

Sample manager flow through needle
Table of Contents

Advertisement

Figure 2–6: Inserting connector into connector port
Connector port
Connector
2.
Using the flat-blade screwdriver, attach the positive and negative leads of the signal cable
to the connector.
Tip:
Refer to the cable-connection label affixed to the rear panel of the module.
Figure 2–7: Positive and negative lead connections
Screw
Connector
Signal cable
3.
Fit the grounding cable's fork terminal on the rear-panel grounding stud and secure the
terminal using the locking nut.
Note:
Use the 9/32-inch nut driver to tighten the locking nut until the terminal does not
move.
June 11, 2019, 715005051 Rev. D
Page 32