2.9.1 Installing the optional lower leak sensor

Warning:
working with hazardous materials. Consult the Material Safety Data Sheets regarding
the solvents you use. Additionally, consult the safety representative for your
organization regarding its protocols for handling such materials.
Warning:
resistant, powder-free gloves when performing this procedure.
Warning:
Required materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• Leak sensor
To install the optional lower leak sensor:
1.
Power-off the sample manager.
2.
Open the fluidics compartment door.
3.
Carefully unpack the new leak sensor.
4.
Align the leak sensor's T-bar with the slot in the side of the leak sensor reservoir and slide
the leak sensor into place.
Observe Good Laboratory Practice (GLP) at all times, particularly when
To avoid personal contamination with toxic materials, wear clean, chemical-
To avoid eye injury, use eye protection when performing this procedure.
Requirement:
Wear clean, chemical-resistant, powder-free gloves when
installing the lower leak sensor.
June 11, 2019, 715005051 Rev. D
Page 34