AMI Codes-II
Maintenance
6.6.
A-96.250.571 / 050517

Cleaning the protective Filter

Switch off the instrument according to instructions in
tion for Maintenance, p. 43
Normally the filter in your sample supply line will retain most debris.
If the filter shows deposits, proceed as follows:
1 Close the main tap of the sample inlet.
2 Close flow regulating valve [B].
3 Unscrew and remove the filter vessel [E] from the
flow cell block [A].
4 Hold the filter [D] on the shaft [C] and unscrew and remove it.
5 Backwash the filter under pressure of tap water.
6 Clean the outside of the filter.
7 Install the filter and the filter vessel again.
8 Establish the sample flow.
9 Adjust sample flow with the regulating valve.
A
Flow cell block
A
B
Flow regulating valve
B
C
Filter shaft
D
Filter
E
Filter vessel
C
D
E
Stop of Opera-
53

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.