Cleaning The Protective Filter - Swann AMI Phosphate-II Operator's Manual

Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
6.7.
44

Cleaning the protective Filter

Switch off the instrument according to instructions in
tion for Maintenance, p.
Normally the filter in your sample supply line will retain most debris. If
the filter shows deposits, proceed as follows:
1 Close the main tap of the sample inlet.
2 Close flow regulating valve [B].
3 Unscrew and remove the filter vessel [E] from the
flow cell block [A].
4 Hold the filter [D] on the shaft [C] and unscrew and remove it.
5 Backwash the filter under pressure of tap water.
6 Clean the outside of the filter.
7 Install the filter and the filter vessel again.
8 Establish the sample flow.
9 Adjust sample flow with the regulating valve.
37.
A
Flow cell block
A
B
Flow regulating valve
B
C
Filter shaft
D
Filter
E
Filter vessel
C
D
E
Stop of Opera-
A-96.250.641 / 290121

Advertisement

Table of Contents
loading

Table of Contents