Setting Error Processing

3-32
The MMS-3 provides Error Processing that enables you to specify
how the recorders and alarms will respond to low and high range
errors.
Low and high range errors occur when an input signal that is within
the capacity of the analyzer exceeds the calibration range of the
probe. The MMS-3 displays an UnderRng message for low range
errors and an OverRng message for high range error. Refer to Range
Error Description in Chapter 2 of the Service Manual for more details
on range errors.
The MMS-3 outputs will also respond to signal and calibration errors.
When a signal error occurs, the outputs will respond in the same
manner as the high range error has been programmed. When a
calibration error occurs, the outputs will respond in the same manner
as the low range error has been programmed. For more details on
signal and calibration errors, refer to Signal Error Description and
Calibration Error Description in Chapter 2 of the Service Manual.
You can program separate actions for low and high range errors for
the alarm relays and recorder outputs. The actions are independently
programmable for each channel.
[PROG]
Press the
key to enter the user program.
Enter Passcode: XXXX
Note: If you have already entered the user program, see Figure 3-2
on page 3-48 to navigate to the System Menu.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
key to select the desired channel.
Programming Menu
[SYSTEM]
AUTOCAL
.
Enter the passcode
1
Use the arrow keys to move the
brackets to SYSTEM and press
[YES].
Advanced Programming
[CHAN]