Repair and Replacement Procedures
Information in This Chapter
This chapter contains procedures for removing and replacing the major
assemblies of your Keysight Technologies N5292A test set controller.
IMPORTANT! Before you begin any testing on a System or instrument,
always inspect the connectors on the instrument, the calibration kit,
cables and adapters.
The N5292A contains extremely sensitive components that can be ruined if
mishandled. Follow instructions carefully when making cable connections,
especially wire harness connections.
The person performing the work accepts responsibility for the full cost of
the repair or replacement of damaged components.
IMPORTANT! For N5224/5B, N5244/5B, and N5267B PNAs When
connecting the 5292-20014 semirigid cable it is critical that the 2.4 mm
end is connected to the PNA's REFn SOURCE OUT port and the SMA end is
connected to the N5292A test set's LO Band connector. Refer to
on page
There are no replaceable parts in the N5293A or N5295A millimeter-wave
heads. If there is a problem with your N5293A or N5295A, return them to
factory for repair. Refer to
page
For network analyzer repair and replacement procedures, see the "Repair
and Replacement" chapter in the service guide for your PNA. Refer to
www.keysight.com.