Keysight Technologies N5290/1A PNA Series Service Manual page 73

2-port and 4-port microwave network analyzer system (900 hz - 110 ghz / 900 hz - 120 ghz)
Table of Contents

Advertisement

System Tests and Adjustments
N5292A Functional Verification Procedure
Figure 3-25
Low Band (LFE) Eterm Test Connections: PNA port 1 connects to the ET LO
(Low) Band. PNA port 2 (bottom) connects to the Test Set NA Port.
Figure 3-26
Low Band (LFE) Eterm Test Connections: Verifies ~60 ohms with DVM from
ground to center conductor.
Keysight N5290/1A Service Guide
3-37