Keysight Technologies N5290/1A PNA Series Service Manual page 53

2-port and 4-port microwave network analyzer system (900 hz - 110 ghz / 900 hz - 120 ghz)
Table of Contents

Advertisement

System Tests and Adjustments
N5292A Functional Verification Procedure
— LO power verification for LOPA/LODA/Phase compensation/Test set mother
— Test set cabling paths front panel connectors to Test set mother board
This section contains the following:
Accessories Used in the N5292A Test Set Operator's Check
Equipment Type
Calibration and Verification Tools
1.0 mm calibration kit
ECal module
Power supply load ET
LO Cal and RF Eterm ET
WBIF and LFE Eterm ET
Cables
System cable
Cables (x6)
3.5 mm (m) to 3.5 mm (m)
GPIB cable
Adapters
1.00 mm (f) to 1.00 mm (f)
SMA (m) to SMA (f) elbow
SMA (m) to SMA (f) thru
a. Use x3 for front panel wideband IF and LFE Eterms.
Keysight N5290/1A Service Guide
board (TSMB) PCA assemblies
(TSMB)
"Accessories Used in the N5292A Test Set Operator's Check" on page 3-17
"Required Equipment" on page 3-18
"Power Sensor and Power Meter Preparation (Required for LO Verify Only)"
on page 3-22
"PNA and Test Set Preparation" on page 3-22
"PNA Calibration (Pna Cal) and Loading the N5292A Test Set Functional
Verification Software (Launch Test)" on page 3-22
Part Number
85059B
N4691B
ET57265
ET57259–1
ET57259–2
N4011-21002
N5240-60092
Z5623A-K20
10833B
Return Loss:
Return Loss:
Return Loss:
a
b
c
14 dB
14 dB
14 dB
3-17