Post-Repair Procedures

Post-Repair Procedures
Post-Repair Procedures
Table 5-1
performed after the replacement of an assembly. These are the recommended
minimum procedures to ensure that the replacement is successfully
completed.
Table 5-1
Post-Repair Procedures (Option 3L5)
Replaced
Required Adjustments
Assembly or Part
Correction Constants (CC)
A1 Source
Perform the following required
Module
adjustments using "A1 Source Module" in
Spot Adjustment of the program.
104
Post Repair Procedures lists the required procedures that must be
Verification
"To Execute the Diagnostic Test" on
page 33
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
RF Output Level Linearity Test
Trace Noise Test
Crosstalk & System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
LF Output Level Accuracy Test (Opt 3L5)
LF Output Level Linearity Test (Opt 3L5)
Input Impedance Test (Opt 3L5)
Absolute Amplitude Accuracy Test (Opt 3L5)
Magnitude Ratio/Phase Frequency
Response Test (Opt 3L5)
Trace Noise Test (Opt 3L5)
Crosstalk and Noise Level Test (Opt 3L5)
Magnitude Ratio/Phase Dynamic Accuracy
Test (Opt 3L5)
Keysight E5061B Network Analyzers