System Tests and Adjustments
N5292A Functional Verification Procedure
Procedure:
1. On the PNA: Press Service > Service Menu > Adjustments > EE Adjust >
2. On the N5292A: Connect LO Fixture (N5292-60127) to port 1.
3. Follow the prompts.
4. Repeat steps
5. Compare results to typical values shown in to
Figure 3-15
Typical LO maximum power (without path corrections)
Keysight N5290/1A Service Guide
LO Drive
OR Run N5292A Launch Test software and click Insertion Loss/Match.
1
through
and
Figure 3-15 on page
3
for the other ports, by clicking LO Verify.
3-25.
Figure 3-15 on page 3-25
3-25