Repair and Replacement Procedures

Removing the N5292A Option 200/400 from the System

Removing the N5292A Option 200/400 from the System
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— 5/16-in torque wrench (set to 8 in-lbs)
— 5/16-in torque wrench (set to 10 in-lbs)
— 5/16-in torque wrench (set to 21 in-lbs)
— 14 mm open-end wrench (set to 4 in-lbs)
— 20 mm open-end wrench (set to 8 in-lbs)
IMPORTANT!
The N52xxB interconnect kits contain 2.4 mm to 3.5 mm cables. It is
important to connect the 2.4 mm end to the PNA and to connect the
3.5 mm end to the N5292A test set controller. Improper connections will
result in damage to your instruments.
It is important to use two wrenches when disconnecting or connecting the
semirigid cables to the 1/2/3/4 port 1.85 mm outputs front ports of the
PNA to avoid stressing the connectors and causing damage to your PNA
(i.e., the 2.4 mm cable end is compatible with the 1.85 mm connector of
the PNA).
Figure 7-2
To avoid damaging your RF ports 1/2/3/4, always use two wrenches
7-6
Keysight N5290/1A Service Guide