Keysight Technologies N5290/1A PNA Series Service Manual page 58

2-port and 4-port microwave network analyzer system (900 hz - 110 ghz / 900 hz - 120 ghz)
Table of Contents

Advertisement

System Tests and Adjustments
N5292A Functional Verification Procedure
— Three SMA or 3.5 mm adapters (female to female)
— Voltmeter
— Power sensor and compatible power meter. The power sensor must be
capable of measuring from -10 dBm to +15 dBm at a frequency of 8 GHz. A
E4413A power sensor may be used.
Power Sensor and Power Meter Preparation (Required for LO Verify Only)
The power sensor and power meter must be prepared to make measurements
at 8 GHz. This may include calibrating the power sensor with the meter and
entering a calibration factor for 8 GHz.
Equipment required:
— 26.5 GHz PNA-B with Option 020
— Power sensor and power meter (with calibration factors for 8 GHz)
PNA and Test Set Preparation
Equipment setup:
1. Connect the Test Set 1/0 cable from the PNA to the Test Set.
2. Connect cables from PNA rear panel IF Out to Test set IF IN.
3. Turn on the PNA and the Test Set.
4. The PNA must be in "Standard PNA" mode.To check the mode, select
Utility > System > Configure > Millimeter Module Configuration. The
Millimeter Module Configuration window will be displayed. "Standard
PNA" should be the Selected Configuration. If not, click on "Standard
PNA" in the list of Available Configuration(s) then click OK.
PNA Calibration (Pna Cal) and Loading the N5292A Test Set Functional
Verification Software (Launch Test)
STOP! Before you begin this section, contact Keysight for information on
downloading the N5292A Test Set Verification software (
Refer to
to
"PNA Calibration (Pna Cal) and Loading the N5292A Test Set Functional
Verification Software (Launch Test)" on page
This procedure prepares the PNA for test set IF/RF insertion loss/Match path
measurements. This procedure also includes how to download the Launch Test
application to a PC. Refer to
3-22
http://na.support.keysight.com/pna/files/N5292A_TESTS.exe
Figure 3-14 on page
3-22.
3-23.