Keysight X Series Measurement Manual page 32

Signal analyzers n9083a & w9083a multi-standard radio measurement application
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Making MSR Measurements
MSR Downlink Signal Measurement
Step
Figure 2-15
MSR Downlink Spectrum Emission Mask Measurement Result
4. Setup the limit.
32
Action
Meas Setup
Offset/Limit,
Press
,
More, Limits then enter the
limit value for each offset
Notes
The Lower or Upper ΔLim result is the
minimum margin from limit line which is
decided by Fail Mask setting. There are four
.
settings for Fail Mask: Absolute, Relative, Abs
AND Rel, Abs OR Rel.
— For Absolute mask, the Lower or
Upper Lim is compared with the
Absolute limit line.
— For Relative mask, the Lower or
Upper Lim is compared with the
Relative limit line.
— For Abs AND Rel mask, the Lower or
Upper Lim is compared with the
higher limit line.
— For Abs OR Rel mask, the Lower or
Upper Lim is compared with the
lower limit line.