Keysight X Series Measurement Manual page 33

Signal analyzers n9083a & w9083a multi-standard radio measurement application
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Making MSR Measurements
MSR Downlink Signal Measurement
Step
5. Select the desired
offset pairs.
6. Select the view and
measurement result.
Figure 2-16
MSR Downlink Spectrum Emission Mask Measurement Result - Carrier Info View
Keysight N9083A & W9083A Measurement Guide
Action
Meas Setup
Offset/Limit,
Press
,
Select Offset
. Select the offset you
want to turn on and press Start Freq,
On. You can change the Start Freq,
Stop Freq and other values for the
offset.
View/Display
Press
and select the
desired view.
Notes
When there are many offsets to measure, you
can increase the Res BW under Meas Setup,
Offset/Limit to increase the measurement
speed.
For views of Absolute Peak Power &
Frequency, Relative Peak Power & Frequency
and Integrated Power, you can use three
measurement types (select using Measure
Setup, Meas Type): Total Power Reference,
PSD Reference and Spectrum Peak
Reference. For Carrier Info view, it contains
the spectrum trace and the carriers
Figure 2-16
information table, see
choose to display offset or absolute frequency
under the menu Carrier Info.
. You can