Making MSR Measurements
MSR Downlink Signal Measurement
Step
Figure 2-14
MSR Downlink ACP Measurement Result - Carrier Info View
5. (Optional) Adjust the
dynamic range.
TIP
For the most accurate ACP measurement results, you maybe able to optimize the level of the
signal measured by the analyzer. Adjust the input attenuation using the Up/Down keys, while
watching the ACP levels shown at the offsets to see if the measurement results improve with
another setting.
6. (Optional) Change
measurement
parameters from their
default condition.
If you have a problem, and get an error message, see the guide "Instrument
Messages", which is provided on the Documentation CD ROM, and in the
instrument here:
C:\Program Files\Keysight\SignalAnalysis\Infrastructure\Help\bookfiles.
30
Action
AMPTD
Press
and adjust the
Attenuation
to 0 dB.
Meas Setup
Press
to see the keys
that are available to change.
Notes
This allows greater dynamic range for this
level of input signal.
Offset/Limits to configure the
Press
settings for each offset.
value for parameters are defined according to
the standard.
Keysight N9083A & W9083A Measurement Guide
The default