Making MSR Measurements
MSR Downlink Signal Measurement
Step
Figure 2-13
MSR Downlink ACP Measurement Result - Result Trace View
4. Select Carrier Info
view.
Noise Correction can reduce the noise contribution of the analyzer to the measurement results as
much as 10 dB. When the measured power is close to the noise floor, turning on the Noise Correction
under the Meas Setup menu can make the measurement more accurate. The correction will only be
valid for current measurement parameters.
To correctly use the Noise Correction feature, you MUST re-calibrate the correction (set to Off,
then On) after ANY measurement parameters are changed. Failure to re-calibrate the Noise
Correction will provide invalid data. When Noise Correction is On, the screen annotation NCORR is
shown below the Input.
Keysight N9083A & W9083A Measurement Guide
Action
View/Display
Carrier
Press
,
Info
. This view contains the spectrum
trace and the carriers information table,
Figure 2-10
see
. You can choose to
display offset or absolute frequency
under the menu Carrier Info.
Notes
Noise Correction
can reduce the noise
contribution of the analyzer to the
measurement results as much as 10 dB.
When the measured power is close to the
noise floor, turning on the Noise Correction
under the Meas Setup menu can make the
measurement more accurate.The correction
will be valid for only the current measurement
parameters.
29