5
Troubleshooting and Error Handling
Well-plate Handler Test
Each movement of the Well-plate Handler can be carried out under manual
control. This is useful during troubleshooting, where close observation of each
of the steps is required to confirm a specific failure mode or verify successful
completion of a repair. Each step command consists of a series of individual
commands which move the components to predefined positions enabling the
specific step to be done.
Testing with the ChemStation
In the ChemStation, the Well-plate Handler test is carried out using the
Well-plate Handler Test dialog box, where the movements are separated into
single-step instructions and multi-step functions (see
74
Figure
53).
1100 Series Sample Capacity Extension